You have no items in your shopping cart
Circulation cooler RC 5 Basic
Powerful recirculating chiller, designed for external cooling applications up to -30°C.
The energy-efficient chiller is operated with an eco-friendly refrigerant and reaches a high power density. This results in a comparatively small working space in the lab.
The speed-controlled PEEK pump allows an indirect, continuous adjustment of the maximum pressure and flow rate. Due to its excellent temperature stability of about ± 0,1 K, the RC 5 can handle even sophisticated applications. The isolated, high-quality 7l bath has an integrated hopper and drain valve, ensuring a safe and clean handling of the thermofluid. A hose can be connected to empty the bath in order to avoid direct contact with the thermofluid. The easy-to-open front flap allows an easy cleaning of the cooling unit’s air filter.
Key Features
- Speed-controlled refrigeration system with 1400 W cooling capacity (@20°C)
- Natural thermofluid R290
- Designed for cooling applications up to -30°C - RT
- Extended operating temperature range until 80°C
- Excellent temperature stability of about ± 0,1K (@-10°C)
- Speed-controlled pump made of PEEK: (0,6 bar; 31l/min)
- Digital level indicator
Interfaces:
- RS 232 and USB
Scope of delivery
- RC 5 basic
- Hose olive NW 8 (2x)
- Hose olive NW 12 (2x)
- USB Cable - USB A to Micro-B, 2 m
| Appliance type | Recirculating chiller |
| Cooling agent | R290 |
| Cooling agent quantity | 90 g |
| Cooling agent pressure max. | 21 bar |
| Cooling capacity (@20°C) | 1400 W |
| Cooling capacity (@10°C) | 1200 W |
| Cooling capacity (@0°C) | 950 W |
| Cooling capacity (@-10°C) | 650 W |
| Cooling capacity (@-20°C) | 450 W |
| Cooling capacity (@-30°C) | 200 W |
| Working temperature | -30 - room temp. °C |
| Operating temperature min. | -30 °C |
| Operating temperature max. (with external heating) | 80 °C |
| Temperature display | yes |
| Temperature control | PT 100 |
| Working temperature sensor | PT 100 |
| Working temperature display | LED |
| Temperature stability DIN 12876 | 0.2 ±K |
| Display resolution | 0.1 K |
| Warning function optical | yes |
| Warning function acoustic | yes |
| Warning function excess temperature | yes |
| Sub-level protection | yes |
| Filling volume | 5.0 - 7.0 l |
| Pump type | Pressure- / suction pump |
| Pump capacity adjustable | yes |
| Pump pressure max. (0 liters discharge flow) | 0.61 bar |
| Pressure pump (suction side) (0 liter flow) | 0.45 bar |
| Flow rate max. (0 bar back pressure) | 31 l/min |
| Pump connection | M16x1 |
| Calibration option | yes |
| Technical data complies with the standard | DIN 12876 |
| Permissible ON time | 100 % |
| Dimensions (W x H x D) | 310 x 546 x 490 mm |
| Weight | 37.5 kg |
| Permissible ambient temperature | 5 - 32 °C |
| Permissible relative humidity | 80 % |
| Protection class according to DIN EN 60529 | IP 21 |
| RS 232 interface | yes |
| USB interface | yes |
| Voltage | 230 / 100 - 115 V |
| Frequency | 50/60 Hz |
| Power input | 1100 W |